Item |
Specification |
1. X-ray High voltage generator |
1.1 |
standard output power: 32kW max output power :40kW |
1.2 |
mA range:10-400mA |
1.3 |
Tube voltage range:40-150kV |
1.4 |
mAs range :0.1-500mAs |
1.5 |
Load time range:0.001s~6.3s |
1.6 |
Input voltage:336VDC±10% |
1.7 |
High voltage transformer structure: The power frequency |
2. X-ray tube |
2.1 |
Focus range:0.6/1.2mm |
2.2 |
Anode heat capacity:300kHu |
2.3 |
Pipe sleeve heat capacity: 1250kHu/900kJ |
2.4 |
The maximum voltage:150kV |
2.4 |
The anode input power: Big focus:50kW, Small focus:22kW |
2.5 |
Anode angle 12° |
2.6 |
The inherent filtration:1.3mm Al/75kV |
3. Collimator |
3.1 |
The inherent filtration:1.2 mm Al /70 kV; |
3.2 |
The collimator is equipped with LED lights(Bright 30s are automatically extinguished) |
3.3 |
The collimator is built into the measuring ruler |
4. The tube bracket structure |
4.1 |
X-ray tube support way: cross arm |
4.2 |
Tube focus distance to the ground:570㎜~1800㎜,±5% |
4.3 |
Telescopic Arm Length Range:710㎜~1240㎜ |
4.4 |
Column Rotation Range:±270° |
4.5 |
Tube Tilting Angle:±180° |
4.6 |
Tube Vertical Rotation Angle:-30°~+90° |
4.7 |
Collimator axial rotation:±90° |
5. Detector |
5.1 |
Detector: PaxScan 4336W |
5.2 |
WIFI connection, Amorphous Silicon, whole panel technology |
5.3 |
Detector valid exposure size:427×344mm |
5.4 |
Pixel size:150 µm |
5.5 |
The effective camera size of the detector:14×17inch |
5.6 |
Gathering matrix:2304×2800 pixels; |
5.7 |
Spatial resolution: 3.4 lp/mm |
5.8 |
Pixel gray level:≥14bit |
5.9 |
Detector cooling mode: natural cooling |
5.10 |
Imaging time:≤5s |
6. Digital image workstation MOC(V02.01.00) |
6.1 |
CPU frequency: CORE i5 |
6.2 |
Memory capacity:4GB |
6.3 |
Hard drive capacity:500 GB |
6.4 |
Display size: 19-inch LCD touchscreen |
6.5 |
Display screen Maximum resolution≥1280×1024; |
6.6 |
64-bit Windows operation system |
7. Image acquisition software |
7.1 |
Patient registration: acquisition, finding the patient data, preparing for the photography |
7.2 |
Radiographic examination: setting system parameters, exposure photography, and getting the image |
7.3 |
Images display, image analysis, processing, and output |
7.4 |
Data management: managing patient data, leading in, leading out, batch files, etc in accordance with DICOM3.0 |
7.5 |
Configure calibration: the system components such as flat-panel detectors, such as tubes for automatic configuration and calibration |
7.6 |
System diagnosis: monitoring system status, diagnosis of the system fault |
8. Battery |
8.1 |
Power supply way |
8.1.1 |
storage battery supply |
8.1.2 |
Battery capacity: Maximum 240 exposures, each time 100mAs, chest image dosage (exposure conditions: 120 kv, 100 mA, 20ms ) |
8.2 |
There is no wall charge when exposed |
8.3 |
Can support hand switch and remote control |
9. The machine parameters |
9.1 |
Mobile mode: infinitely variable speed |
9.2 |
Working mode: intermittent loading, continuous operation |
9.3 |
Appearance size |
9.3.1 |
Whole machine width:670 mm |
9.3.2 |
Whole machine length::1235 mm |
9.3.3 |
Whole machine height:1825 mm |
9.4 |
Whole weight:518kg |
9.5 |
Drive mode: dual motor drive, respectively control left and right driving wheel |
9.6 |
Collision prevention: the obstacle contact sensor |
10. External interface |
10.1 |
Dual display interface (VGA/DVI) |
10.2 |
Wireless network data transmission(WIFI) |
10.3 |
Cable network interface |
10.4 |
USB(4units) |
10.5 |
Choose a chest rack |
10.6 |
The selection of moving flatbed |
11. Aftersale service |
11.1 |
The machine warranty 1 year |
11.2 |
1 set of complete operation and maintenance manual |
11.3 |
Provide site application and service training |
11.4 |
Free service hotline |
12. copyright certificate |
12.1 |
DR service control application software |
12.2 |
mobile Dr embedded system software |
12.3 |
mobile Dr generator control software |
13. Related patents |
13.1 |
Utility model patent certificate – a mobile digital medical X-ray photogram-system column rotating mechanism |
13.2 |
Utility model patent certificate – a mobile digital medical X-ray system anti-collision mechanism |
13.3 |
Utility model patent certificate – a mobile X-ray photographic system telescopic arm mechanism |