Specifications:
● Power requirement:AC220V-230V/50Hz or AC110-120V/50-60Hz, 1300VA
● Control Mode: air mode &baby mode controlled by micro-computer
● Air Temperature Control Range:25°C-37°C (override mode 37-39°C)
● Infant Temperature Control Range:34-37°C (override mode 37-38°C)
● Accuracy of Skin Temperature Sensor: ±0.2°C
● The temperature difference between displayed incubator temperature and average incubator
temperature (under steady temperature condition): ≤0.5°C
● Temperature Uniformity: ≤0.8°C(when mattress is in horizontal position)
● Failure Alarms: over-temperature alarm, temperature deviation alarm, sensor alarm, motorized fan failed alarm,
power failure alarm, sensor box position alarm, tank position alarm, system alarm and many others.
● Humidity Display Range: 0%RH~99%RH
● Humidity Control Range: 0%RH~90%RH
● Accuracy of Humidity Display:±5%RH (when environment temperature is 25oC, environment humidity is45%RH±5%RH)
● Inclination of Bassinet: ±8°
● Noise Level within Canopy: ≤45dB(A)
● Indication Range of Oxygen Concentration: 0%~99%
● Control Range of Oxygen Concentration: 20%~60%
● Accuracy of Oxygen Display Concentration: ±2% (when oxygen concentration is set ≤25%),
±3% (when oxygen concentration is set >25%)
● Accuracy of Oxygen Control Concentration: ±4%
● Weight Range:100-8000g (scale)
● Resolution for weight indication: 1g increments (scale)
● Accuracy of Scale: ±1% (scale)
Standard Configuration:
● Infant Compartment
● Main Unit: Vertical Height Adjustment (VHA) stand, controller, IV stand and shelf
Optional:
● Scale, monitor shelf and disposable skin temperature sensor
Reviews